For nearly a decade, the NV® Microlaser has set the bar for convenience, portability and ease of use among all dental soft-tissue diode lasers. The next-generation NV® PRO3 Microlaser continues that award-winning tradition. Optimized for all of your periodontal, restorative and orthodontic procedural needs, the latest evolution in cordless soft-tissue lasers enables you to deliver the benefits of laser dentistry to each patient, while increasing practice production across all departments.

Description

The Best Just Got Better

Introducing the new NV® PRO3 Microlaser

NV® PRO3

For nearly a decade, the NV® Microlaser has set the bar for convenience, portability and ease of use among all dental soft-tissue diode lasers. The next-generation NV® PRO3 Microlaser continues that award-winning tradition. Optimized for all of your periodontal, restorative and orthodontic procedural needs, the latest evolution in cordless soft-tissue lasers enables you to deliver the benefits of laser dentistry to each patient, while increasing practice production across all departments.

 

Indications for Use

 

Procedure Technique Tip Mode Watts
Apthous Ulcer Non-contact Non-initiated Pulse 1.4 W
Bacterial Decontamination Non-contact Non-initiated Continuous 0.8-1.0 W
Biopsy Contact Initiated Continuous 1.2 W
Biostimulation Contact Initiated Continuous 1.0 W
Class V Restoration Contact Initiated Continuous 1.0 W
Cuspid Exposure Contact Initiated Continuous 1.0 W
Distal Wedge Contact Initiated Continuous 1.5 W
Draining Abscesses Contact Initiated Continuous 1.2 W
Fibroma Contact Initiated Continuous 1.2 W
Frenectomy Contact Initiated Continuous 1.2 W
Gingivectomy Non-contact Initiated Continuous 1.0 W
Gingivoplasty Contact Initiated Continuous 1.0 W
Hemostasis Contact Initiated Continuous 1.5 W
Hypertrophic Tissue Contact Initiated Continuous 1.5 W
Hyperplasia Contact Initiated Continuous 1.2 W
Implant Recovery Contact Initiated Continuous 1.2 W
Laser Troughing Contact Initiated Continuous 1.0 W
Operculectomy Contact Initiated Continuous 1.5 W
Ovate Pontic Contact Initiated Continuous 1.2 W
Recontouring Contact Initiated Continuous 1.0 W
Sulcular Debridement Non-contact Non-initiated Continuous 0.8 W
Tissue Tag Contact Initiated Continuous 1.2 W
Tongue Tie Contact Initiated Continuous 1.2 W

 

Click here to download Diode Laser Research pdf

 

Additional information

Product Specifications

• Dimensions of Base Charger: 2.7” H x 2.9” W x 8.0” L
• Dimensions of Laser: 6.9” L x 0.65” diameter
• Weight: 1.9 ounces
• Laser classification: Class IV laser device
• Delivery system: Optical Fiber
• Wavelength: Laser 808 nm ± 5 nm
• Maximum power: 2 Watts ± 20%
• Aiming beam wavelength: 650 ± 10 nm
• Aiming beam power: 0.5 mW max
• Beam divergence: 617 mrad
• Power range: 0.1 watt to 2.0 watts
• Pulse frequency: Fixed 10 Hz
• Pulse duration: Fixed 0.05 seconds
• Duty cycle: Pulse Mode: 50%
• Continuous Wave Mode: 100%
• Audible notification: Yes
• Power requirements: 100-240 VAC @ 50 to 60 Hz
• Charger current: 0.6 Amps
• Battery: Rechargeable lithium-ion
• 20 minutes continuous lasing time @ 1.2 watts: 8 hours standby time
• Wireless Foot Pedal frequency: 2.4 GHz
• Foot Pedal power source: AA Batteries (provided)

Kit Contents

• NV PRO3 Microlaser Main Body
• Lithium-Ion Batteries (2)
• Pairs of Safety Glasses (3)
• Base Charger
• Power Supply for the Charger
• Wireless Foot Pedal
• AA Batteries for the Foot Pedal (2)
• Disposable Tips — 5 mm (5); 7 mm (5)
• Warranty Information Card
• DANGER Warning Sign
• User Manual
• Pack of Carbon Film

Reviews

There are no reviews yet.

Be the first to review “NV® PRO3 Microlaser”

Your email address will not be published. Required fields are marked *

preloader